AI Troubleshooting

Diagnose LC-MS chromatography and ionization problems

M

Symptom-Based Diagnosis

Pick what you're observing on the instrument

Likely Causes

  • Secondary silanol interactions with basic analytes on older/legacy silica
  • Column void or bed collapse
  • Extra-column volume / dead volume in tubing or fittings
  • Injection solvent stronger than mobile phase (solvent mismatch)

Recommended Fixes

  • Switch to a high-purity, end-capped BEH/HSS-type phase rated for basic compounds
  • Add a small amount of a silanol-blocking additive (e.g. low mM ammonium formate) at slightly acidic pH
  • Reduce injection volume or dilute the sample in weak solvent (match to mobile phase A)
  • Replace the column or check for a void by looking at a well-retained test-mix peak shape