Symptom-Based Diagnosis
Pick what you're observing on the instrument
Likely Causes
- Secondary silanol interactions with basic analytes on older/legacy silica
- Column void or bed collapse
- Extra-column volume / dead volume in tubing or fittings
- Injection solvent stronger than mobile phase (solvent mismatch)
Recommended Fixes
- Switch to a high-purity, end-capped BEH/HSS-type phase rated for basic compounds
- Add a small amount of a silanol-blocking additive (e.g. low mM ammonium formate) at slightly acidic pH
- Reduce injection volume or dilute the sample in weak solvent (match to mobile phase A)
- Replace the column or check for a void by looking at a well-retained test-mix peak shape