Practical Pharma Science™
Deeper courses, SOP kits, and validation templates from the makers of LC-MS Assist Pro.
Shop on PayhipLC-MS Reference Articles
ESI vs. APCI: choosing the right ionization source
ESI suits polar-to-moderately-nonpolar, thermally labile, and larger analytes across a wide MW range. APCI works better for smaller, less polar/neutral molecules and is more tolerant of higher-salt or less-volatile mobile phases. Start with ESI for typical drug-like small molecules; switch to APCI if ESI gives poor/no response for a neutral, low-polarity analyte.
MS-compatible mobile-phase additives — quick reference
Volatile: ammonium formate/acetate (2–10 mM), formic/acetic acid (0.05–0.2%), ammonium bicarbonate (basic-mode negative ESI). Avoid: phosphate, borate, TRIS, and non-volatile salts — they suppress ionization and foul the source. TFA is MS-compatible but strongly suppresses positive-mode signal; use sparingly.
Building an MRM method from scratch
1) Full-scan MS1 to confirm the dominant precursor adduct. 2) Product-ion scan with a CE ramp to find 2–3 diagnostic fragments. 3) Optimize CE per transition. 4) Pick the most intense, most selective fragment as quantifier and a second as qualifier. 5) Verify the ion ratio (quantifier:qualifier) stays consistent across the calibration range per ICH M10.
Assessing matrix effects & ion suppression
Post-column infusion of the analyte while injecting blank matrix reveals suppression/enhancement zones across the chromatogram — retention should avoid the void-volume suppression zone. For quantitation, matrix-matched calibration or a stable-isotope-labeled internal standard is the most robust way to correct for matrix effects.
ICH M10 bioanalytical validation — key acceptance criteria
Accuracy/precision within ±15% (±20% at LLOQ). Calibration curve with ≥6 non-zero levels. Selectivity in ≥6 individual matrix lots. Matrix effect assessed via post-extraction spike / IS-normalized matrix factor CV ≤15%. Carryover ≤20% of LLOQ response (≤5% of IS). Dilution integrity and stability (bench-top, freeze-thaw, long-term) must be demonstrated.
When a C18 isn't enough: orthogonal column chemistries
Phenyl-hexyl and PFP phases add pi-pi and dipole selectivity useful for separating isobaric interferences, halogenated compounds, and positional isomers that co-elute on a standard C18. HILIC (amide/BEH) is the go-to for very polar analytes (logP < -1) reversed-phase can't retain.